| |
|
|
Basic Measurements
The 931A, 930A, and 929A measure all of the basic quantities:
true-rms voltage and current, frequency and phase angle. PowerDSA™
analysis measures these quantities more accurately than ever before.
Accuracy is 0.05% for voltage and current and 0.05° for phase.
The proprietary PowerDSA™
narrow-band mode even measures the fundamental signal alone,
rejecting the effects of harmonics and noise. In wide-band mode, the
effects of all harmonics and noise are included. Phase angle is
always true, fundamental phase, and frequency is accurate even with
large harmonics causing multiple zero crossings. Accuracy is never
degraded, even with real-world signals. |
 |
|
 |
|
 |
| |
|
|
Power Qualities
The PowerDSA™ instruments measure power quantities, too, with
accuracy unprecedented for a lightweight, portable instrument.
Measurements are made in accordance with IEEE standards, including
the effects of harmonics and reactive power. Watts (W), watthours
(Wh), volt-amperes (VA), volt-ampere hours (VAh), volt-amperes
reactive (VAR), volt-amperes reactive hours (VARh), and power factor
(PF): PowerDSA™ analysis measures them all, with 0.11% basic
accuracy. |
|
|
 |
| |
| |
|
|
Harmonics
Measure harmonics and view the results graphically; as summary
numbers, such as total harmonic distortion (THD), even harmonic
distortion (EHD), odd harmonic distortion (OHD) or K-factor; or as
individual harmonic amplitude and phase. The bandwidth extends to
over 3 kHz for accurate measurement to the 50th harmonic
on 50 or 60 Hz systems, now you can know for sure what is really
happening on your system. |
|
|
 |
| |
| |
|
|
Waveform Display
You can view the signal waveforms for both channels on the high
contrast 240x64 graphic display. Both channels are normalized to
Channel 1 fundamental phase, so you can see, for example, the
relationship between current waveform distortions and voltage phase. |
|
|
 |
| |
| |
|
|
Timer
(Model 931A)
The Model 931A's timer/counter (not available on the Model
930A/929A) measures relay operating times or elapsed test times, or
counts input rates and events. View times in seconds or in terms of
power-line cycles. Both inputs are independently opto-isolated for
safety and flexibility. |
|
|
 |
| |
| |
|
Three Phase
Measurement
The 931A, 930A, and 929A include a full three-phase input section,
for automated three-phase measurement sequences. PowerDSA™
analysis measures two signals at a time, and the results are
combined into four complete three-phase displays.
You can select from the following
three-phase display modes:
- Basic three-phase display
View voltage, current, phase, frequency and power quantities on
one convenient display.
- Vector display
View voltages, currents and phase angles with their vector
representation.
- Voltage/Current Sequence display
View voltage, current and phase along with positive, negative and
zero sequence values.
- Energy display
View voltage, current, phase, frequency and energy quantities on
one convenient display.
As a power trend monitor and recording
system, these PowerDSA™ instruments can verify phase
relation, phase rotation, power direction, load balance and
positive, negative, and zero sequence of voltage and/or current as
well as calibrating and verifying in-service performance of
Disturbance, Fault and Transient Recorders. |
|
 |
 |
 |
 |
| |
| |
|
|
Transducer Calibration
(Model 931A)
The outstanding accuracy of the Model 931A with PowerDSA™ is
ideal for transducer calibration. The Model 931A includes the
necessary dc voltage and current measurements to accurately measure
transducer outputs, using a separate, opto-isolated input section.
With only an external source or load box, or when making in-service
measurements, the Model 931A provides complete field transducer
calibration. |
|
|
 |
| |
| |
|
|
CT/PT Ratios
Correction factors for external CTs and/or PTs, can be entered to
display the measured results in input-side units. You can even
measure ratios using the instrument's CH1/CH2 function. This example
shows a nominal 200:1 (or 1000:5) current ratio; if Channel 1 is a
CT burden voltage and Channel 2 CT secondary current, the result is
the loop resistance in ohms. |
|
|
 |
| |
| |
|
|
Extended Measurement Ranges
For greater accuracy, the Model 09311A Auxiliary CT allows
measurement of signals up to 400 amps with total basic accuracy of
0.1%. This CT mounts directly to the Model 931A, 930A, or 929A
current input connectors and may be used for one two or three of the
current inputs, depending on your needs. |
|
|
 |
| |
|