 Model 931A/930A/929A Power System Analyzer Specifications

Navigation:

 Arbiter Systems Index > 930 Features > 930 measurements > 930 Specifications > 931A Accessories >
   

Model 931A Power System Analyzer Specifications

 
 
Input
     

Basic Inputs & Measurements:
The Arbiter Systems® Model 931A Power System Analyzer has two main measurement channels, Channel 1 and Channel 2. Any voltage or current input signal may be selected for either channel. For basic measurements (voltage, current, frequency, phase angle) any combination of inputs may be used. For power and energy measurements (active power, apparent power, reactive power and power factor), one voltage and one current must be selected. For three-phase measurements, the input configuration is selected automatically, based on the measurement type (for example, 3-phase 4-wire 3-element).
 

  Voltage:  
  Input Range

1.5 to 750 Volts rms (underrange to 200 mV) [2 to 1000 Vdc (Model 931A Only)]

  Inputs

Four; A, B, C, N

   

Phase-to-Phase

   

Phase-to-Neutral

   

Phase-to-Synthesized Neutral (A+B+C / 3)

  Impedance

1 megohm

  Leakage

< 3.5 mA per IEC348 and UL1244

     
  Current:  
  Input Range

 0.04 to 20 Amps rms (underrange to < 1 mA)

  Inputs

Three; A, B, C, plus synthesized neutral

  Burden

0.01 ohm maximum

  Isolation

Transformer, 1000 Vrms

  Neutral

Synthesized, -(A+B+C)

     
  Timer: (Model 931A Only)
  Inputs 

Two; 4 to 300 Vdc; 20 to 300 Vrms; or dry contact/thyristor output

  Isolation

Optical, 300 Vrms, each channel

     
  Transducer: (Model 931A Only)
  Range

0 to 1, 0 to 100 mA dc and 0 to 10 Vdc

  Protection

Overvoltage to 120 V, both inputs

  Isolation

Optical, 300 Vrms

     
     
Measurements
     
  Voltage and Current:
  Method

Wideband: True rms, 3 kHz Bandwidth

   

Narrowband: Fundamental magnitude

  Accuracy

0.05% of reading

  Underrange

< 1% of reading, typical at 0.3 mA rms

  DC voltage

0.1% of reading + 25 mV dc, typical (Model 931A only)

     
  Phase Angle:
  Input

Channel 1 to Channel 2

  Range

0 to 360° or ±180°

  Accuracy

0.05°

  Underrange

< 1°, typical at 0.3 mA rms

     
  Frequency:
  Input

Channel 1

  Range

20 to 500 Hz (underrange to 5 Hz)

  Accuracy

0.005% of reading

     
  Harmonics:
  Input

Channel 1 or Channel 2

  Range

2nd to 50th Harmonic (50 or 60 Hz fundamental)

  Accuracy

5% of THD reading + 0.01%

  Display
 

THD; K-factor; Amplitude bar graph; and individual harmonic magnitude and phase (simultaneous)

     
  Waveform:
 

Display

Channel 1 and/or Channel 2

     
  Power / Energy Quantities:
  Range

0 to 99999 MVA or MVAh

   

±99999 MVAR or MVARh

   

±99999 MW or MWh

   

±1.0000 PF, lead or lag

  Accuracy

0.11% of VA, for VA, VAR, and W, 0.001 PF

Transducer Accuracy

0.05% of reading + 0.01% range  (Model 931A Only)

     
  Timer (Model 931A Only):
  Range

0.0001 to 9999.9 seconds, or 0.01 to 999999 cycles

  Accuracy

0.005% of reading + 1 digit

  AC Trigger

Add 1 ms maximum, 0.15 ms at 120 Vrms

     
Interface
     
  Operator Interface:
  Display

240x64 graphic LCD with cold-cathode fluorescent lamp (CCFL) backlight

  Keyboard

21 function keys plus On/Off

  Memory

EEPROM (calibration data)

   

Battery backup RAM (setup and stored results)

   

Real-time clock

  Data

Instrument calibration data

   

User setups (up to six)

   

Logged data (15 to 200); time-tagged

     
  System Interface (Model 931A and 930A Only):
  RS-232

1200 to 115,200 baud; 7 or 8 data bits; 1 or 2 stop bits; even/odd/no parity

  Isolation

Optical, 300 Vrms

     
Power Requirements
     
  Internal Power:
  Type

Sealed lead-acid battery

  Operation

8 hours typical

  Charge

8 hours typical; fast + float charge

     
  External Power:
  Range

85 to 264 Vac, 47 to 440 Hz, 15 VA maximum

   

110 to 250 Vdc, 15 W maximum

  Safety

Designed to meet UL, CSA, VDE

     
General  
     
  Physical:  
  Size

205mm x 305mm x 225 mm (8" x 12" x 8.75")

  Weight

5.8 kg (12.8 lbs), maximum

     
  Environmental:
Temperature

-10° to +50° C Operating Temperature

   

-40° to +75° C Non-operating Temperature

  Humidity

Non-condensing

     
Options  
  Option Number

Description

  930Aopt01

16 MB memory

  931Aopt01

16 MB memory

     
     
 

Accles & Shelvoke| Aqua-Tronics | Arbiter Systems | Clare Instruments | Haefely | Hipotronics | HV Diagnostics | Megger | Tettex | T & R Equipment  
Home Page | High Current | High Voltage | Cable Fault | Oil Test | Transformer | Insulation 

 
Macey's Electrical Pty Ltd
ABN 17 344 601 468
Unit 1, 49 Gavenlock Road,  Tuggerah,  Sydney  Australia  2259

 

Phone: 61 2 4353 4644 

Mobile: 0412 678 925

Fax: 61 2 4353 4424

mailto:sales@maceyselectrical.com.au
 

Copyright: 1978 to 2007                                                   Last update: Friday, 03 October 2008