 Calibration Test Unit V242
Navigation:  Clare Test Equipment Index > Clare Accessories Index >
 






 

V242 Calibration Check Unit

The V242 Calibration Check Unit is intended for on-site verification of the standard range of Clare test stations and stand-alone core testers. It is recommended that this instrument is returned to a Clare service depot for an annual re-calibration.   

  • Earth Resistance: @ 90, 115, 450, 575 milliohms (calibration limits ±5%).
  • Flash Test: @ 0 to 5000 Volts a.c. and 0 to 2500 Volts d.c. (calibration limits ±1.5%).
  • Insulation Resistance: @ 100K, 200K, 400K, 1M, 2M, 5M, 10M, 20M, 50M and 100Mohm. (calibration limits ±1% except 50M and 100M ±4%) .
    The flash trip value of 5 milliamps can be checked by using the 200K resistance value, which when 1000 volts is applied will induce a current flow of 5mA.

 