 Clare G104 Tester Specifications
Navigation:  Clare Comprehensive Test Station Index > GENIUS/GENIUS LITE Index >
 
Clare G-Series of advanced, microprocessor controlled, comprehensive test stations, designed for production line and type testing requirements. G104 GENIUS Plus
Advanced comprehensive test station plus earth
leakage monitoring

Specifications/Measurement Functions
 

 
 
Phase Neutral Continuity Earth Bond

Test Voltage:

5V D.C.

Test Voltage:

7.5V A.C.

Acceptance Range:

10Ω to 1000Ω

Output Current Range:

5.0A to 25.0A settable in

Accuracy:

±2% ±2Ω

 

0.1A steps

Test Duration:

0.5s to 10.0s

Compliance Ranges:

0Ω to 150MΩ at 25A

Trip Level upper/lower:

10Ω to 1000Ω ±2% ±2Ω

 

0Ω to 500MΩ at 10A
   

Measurement Range MΩ:

0Ω to 1000MΩ ±2% ±1MΩ
AC/DC Flash

Measurement Range mV:

0 to 4000mV ±2% ±1mV

Test Voltage:

0 to 4.0kV in 10V steps

Test Duration:

0.5s to 10.0s
  0 to 5kV settable in

Trip Level upper/lower:

1Ω to 1000MΩ
  10V steps    

Class I output:

0 to 2.5kV DC Insulation Resistance Test

Class II output:

0 to 4kV

Test Voltage:

500V D.C.

Output Rating:

500VA

Trip Range:

2.0 to 500MΩ ±5% ±0.2MΩ

Waveform:

distortion less than 1%

Test Duration:

0.5s to 10.0s

Line/Load Regulation:

better than 1%    

Maximum Slew Rate:

10kV/s Load Test

Ramp Timer (up/down):

0.5s to 10.0s

Supply Voltage:

230V A.C.

Dwell Timer:

0.5s to 300.0s

 

110V A.C.

Leakage Trip Upper:

A.C.: 0.1mA to 10.0mA

 

(Duty Cycle 30%)
  in 0.1mA increments

Measurement Range:

0 to 10.0A
  D.C.:0.1mA to 5.0mA

Test Duration:

0.5s to 300.0s
  in 0.1mA increments

Trip Level upper/lower:

0.1 to 10.0A

Leakage Trip Lower:

A.C.: 0.1mA to 10.0mA

Accuracy:

±2% ±0.05A
  in 0.1mA increments    

*no test detection

D.C.:0.1mA to 5.0mA Earth Leakage Monitor
  in 0.1mA increments

Application:

simultaneous with load test

Trip Accuracy:

±2% of setting

Measurement Range:

0 to 10.0mA
  ±0.5mA

Trip Level upper/lower:

0.1 to 10.0mA

Failure Mode:

breakdown and/or trip

Accuracy:

±2% ±0.05mA

Failure Indication:

visual and audible    
    General Specifications
ARC Detection

Operating Mode:

Automatic or Manual
  detection circuit which

Temperature:

storage: 0°C to +40°C
  prevents damage to   operation: +10°C to +35°C
  device under test

Humidity:

95% max. non-condensing
   

Dimensions:

488mm (L) x 472mm (H)
                         x 475mm (D)
   

Weight:

50kg
       
    Options
   

Alternative Supply:

110V A.C. operation
   

3, 5, or 10 meter Cables:

RS232 cable
   

Printer:

Star industrial label printer
   

High visibility beacon:

low voltage red/green
   

 

surface mounted
   

ENC Enclosures:

Testing enclosures

Accles & Shelvoke| Aqua-Tronics | Arbiter Systems | Clare Instruments | Haefely | Hipotronics | HV Diagnostics | Megger | Tettex | T & R Equipment  
Home Page | High Current | High Voltage | Cable Fault | Oil Test | Transformer | Insulation 

 
Macey's Electrical Pty Ltd
ABN 17 344 601 468
Unit 1, 49 Gavenlock Road,  Tuggerah,  Sydney  Australia  2259

 

Phone: 61 2 4353 4644 

Mobile: 0412 678 925

Fax: 61 2 4353 4424

mailto:sales@maceyselectrical.com.au
 

Copyright: 1978 to 2007                                                   Last update: Friday, 03 October 2008