| Functional
Specifications |
| |
|
| Applied Voltage |
0 to 5kV AC and DC |
| Output Rating |
500VA source |
| Output Adjustment |
0.5 to 5kV in 10 volt
steps |
| Output Compliance Range |
0 to 150 MΩ
@ 25 Amps |
| Waveform AC |
Sinewave, distortion ≤1% |
| Ripple DC |
≤5% at 5kV |
| Line and Load
Regulation |
within 1% |
| Slew Rate (soft start
G503L/M) |
10kV/second |
| Ramp Timer (G503 N/P) |
0.5 to 10 seconds in 0.1
second increments |
| Dwell Timer |
0.5 to 100 seconds in 0.2
second increments |
| Trip Levels - Upper |
AC 0.1 to 20.0mA in 0.1mA
increments |
| |
DC 0.1 to 5.0mA in 0.1mA
increments |
| Trip Levels - Lower and
|
AC 0.1 to 20.0mA in 0.1mA
increments |
| no test detection |
DC 0.1 to 5.0mA in 0.1mA
increments |
| Trip Accuracy |
±2% of setting + 0.05mA
equal to 0.1mA of set value |
| Failure Mode |
Breakdown and/or trip |
| Failure Indication |
Visual and audible |
| Discharge time
|
<10mS 9half cycle) |
|
at failure or end of test |
0.5 seconds (minimum) |
| Arc Detection |
Detection circuit to
prevent arc damage to DUT, instrument and |
| |
for EMC conformance |
|
General Specifications |
| |
|
| Display |
Vacuum Florescent Display
(VDF) 4 lines x 20 characters |
| Data Input |
Front panel keyboard
|
Status Indication
|
Front panel indicators:
- test on
- pass
- fail/fault reset
- test complete
|
|
Function Switches
|
Front panel:
- manual fault reset
- security keylock
|
|
| Test Output Connections |
Two HT output sockets |
| Inputs/Outputs |
4-pin connector (Beacon) |
| |
COMMS 1 (RS232)
|
| |
COMMS 2 (RS232) |
| |
Printer port (RS232) |
| |
External Device port |
| |
3 pin guard socket |
| |
Pneumatic connection |
| Supply |
230 Volt 50Hz (other
options available) |
| Power Consumption |
5 Amp maximum |
| Temperature (storage) |
-20°C
to +50°C |
| Temperature
(operational) |
+5°C
to +40°C |
| Humidity |
0 to 90% RH,
non-condensing |
| Mechanical |
Bench standing, steel case
construction |
| Size |
370mm (L) x 185mm (H) x
450mm (D) |
| Weight |
18kg |