The 100 HVT Series AC test sets are specifically designed to
perform quick and accurate AC dielectric tests. The 100 HVT
has a maximum output of 100 kV at 50 mA or 50 kV at 100 mA.
This test set can be used for testing bucket trucks and aerial
platforms, hot sticks, blankets, ropes or any other high
voltage AC test requirement. They are particularly suitable
for meeting test procedures of the ANSI/SIA A92.2 1990
specification applied to the insulating boom of work platforms
and bucket trucks.
A triple range current meter enables low current readings of 0
to 1/10/100 mA on the 100 HVT. A triple range voltmeter is
connected directly across the output for maximum accuracy. The
100 HVT has a 50 kV tap that is used for bucket liner testing
and other low voltage, high current test requirements.
|
|

|
Features
- Continuously adjustable output
from zero to full voltage.
- Compact design.
- Single end grounded high voltage
test transformer.
- Separate guard and return.
- Triple range current meter.
- Triple range voltmeter.
- Hand cart mounted.
- Zero start interlock.
- External interlock provision.
- Input power circuit breaker.
- Fast acting circuit breaker.
- 50kV tap for bucket liner testing.
|
|
Benefits
- Test to ANSI/SIA A92.2 1990 specifications.
- Low leakage current measurements.
- Guard circuit eliminates stray
leakage
currents.
- Safe and easy operation.
- Portable for use in the field.
- Ideal for any high voltage AC
insulation test requirements.
|
Applications
Ideal for any high voltage AC
insulation test requirement:
- Aerial platform hipoting.
- Bucket truck hipoting.
- Hot stick testing.
- Rotating machinery testing.
- Any high voltage AC test
requirement (within current limitations).
|
|
| Specifications |
| Model |
100HVT |
| Output |
100kV @ 50mA = 5kVA/50kV @
100mA = 5kV |
| Output Current |
0 to 10mA |
| Current
Meter |
Triple range 0 to 1/10/100mA |
|
Voltmeter |
Triple range 0 to 25/50/100kV AC |
| Duty Cycle |
5 minutes on, 15 minutes off, repeated 6
times, then off for 2 hours |
| Output
Termination |
High Voltage Toroid |
| Meter
Accuracy |
±2% of full range |
|
Distortion |
less than 5% of sinusoidal wave |
Cables
|
25ft (7.6 meters) Controller
to HV Tank with Plug Connection |
|
(2) 25ft Shielded Cable for
Return and Voltmeter |
|
(2) 25ft Guard/Ground Cable |
|
10ft (3 meters) Input Power
Cable |
| Dimensions |
17.4" (W) x 14.5" (D) x
31.9"
(H) (441mm (W) x 368mm (D) x 810mm (H) |
| Net Weight
HV Section |
135lb (61kg) |
|
Controls |
62lb (28kg) |
|
Complete Unit |
240lb (109kg) |
| Shipping Weight |
306lb (139kg) |
|
| Note: |
For Input Voltage of 120V/60Hz model number suffix -A
For Input Voltage of 220V/50Hz model number suffix -B |
|
| |