1.
OTS60
Use either ‘manual’ by setting 10 kV, 20 kV......60 kV in turn
or use the slowest ramp rate of 0,5 kV/sec.
2. OTS60 auto
Use either ‘manual’ by setting 10 kV, 20 kV......60 kV in turn
or use the ASTM D1816 which ramps slowly at 0,5 kV/sec.
3. OTS60AF
Use either the calibrate option or the ‘Withstand’ (Proof)
program by setting 10 kV, 20 kV......60 kV in turn or use the
ASTM D1816 which ramps slowly at 0,5 kV/sec.
|
4. OTS75 auto
Use either ‘manual’ by setting 10 kV, 20 kV......75 kV in
turn or use the ASTM D1816 which ramps slowly at 0,5
kV/sec.
5. OTS80AF
Use either the calibrate option or the ‘Withstand’ (Proof)
program by setting 10 kV, 20 kV......80 kV in turn or use the
ASTM D1816 which ramps slowly at 0,5 kV/sec.
6. OTS100AF
Use either the calibrate option or the ‘Withstand’ (Proof)
program by setting 10 kV, 20 kV......100 kV in turn or use the
ASTM D1816 which ramps slowly at 0,5 kV/sec. |