Specifications
|
|
Main Output
The main output on the unit has four taps,
allowing the selection of output voltages up to 240V and output
currents up to 200A. |
| |
|
|
Current |
Output @230V |
|
Range |
Continuous |
5 min* |
1 min* |
6 sec** |
O/CV |
Load V |
|
10V |
33A |
67A |
100A |
200A |
10.6V |
8.7V@100A |
|
35V |
10A |
20A |
30A |
- |
36V |
32V@30A |
|
100V |
3A |
6A |
10A |
- |
108V |
99V@10A |
|
240V |
1A |
2A |
3A |
- |
276V |
259V@3A |
|
240Vdc |
1A |
2A |
3A |
- |
|
|
| Note: |
*Off time of 15 minutes. On times based on an ambient
temperature of 25°C.
**6 second intermittent ratings available with 230V supply. |
| |
Protection: over current trip, duty
cycle trip, thermal monitoring. |
|
|
|
|
Current Limit
Mode
The main output has a current limit
mode that gives very fine control of output currents up to 10A. It
also allows fine current control into very low impedance loads such
as digital relays. |
| |
|
|
Current |
Output @230V |
|
Range |
Short Circuit |
Continuous |
5min |
2min |
O/CV |
Load V |
|
10V |
10A |
3A |
6A |
10A |
8.6V |
5V@6A |
|
35V |
3A |
1A |
2A |
3A |
29V |
132V@30A |
|
100V |
1A |
0.3A |
0.6A |
1A |
88V |
99V@10A |
|
240V |
0.3A |
0.1A |
0.2A |
0.3A |
224V |
259V@3A |
|
| |
|
Auxiliary DC Output |
| |
|
Range |
Maximum A |
Continuous
Rating |
|
12-60V |
1A |
25W |
|
60-220V |
0.23A |
25W |
|
Protection: current limit |
| |
| Phase-shifting AC Output |
| |
|
|
Maximum Output
Voltage |
Current |
Current |
|
Range |
No Load |
Full load |
Continuous |
5 min on/15 min
off |
|
0-130V |
144V |
125V |
0.23A |
0.46A |
|
0-260V |
288V |
250V |
0.11A |
0.23A |
|
0-6V |
6.6V |
5V |
5A |
10A |
|
Frequency range: 45—100Hz
Phase angle: 0—±180°
Protection: current limit and electronic trip. |
| |
Metering
The output is metered by a digital
true RMS system with a single cycle capture memory ammeter—whenever
the timer stops and the output is switched off, the current reading
is held on the display. The RMS value of a single cycle can be
captured. A current trip is set to 110% of full scale of the
selected metering range. |
| |
|
Range |
Resolution |
Trip
current |
Accuracy |
Acquisition
time |
|
5.000A |
0.001A |
5.5A |
±0.5% rdg±5d |
20ms |
|
20.00A |
0.01A |
22A |
±0.5% rdg±5d |
20ms |
|
50.00A |
0.01A |
55A |
±0.5% rdg±5d |
20ms |
|
200.0A |
200.0A |
220A |
±0.5% rdg±5d |
20ms |
|
| |
| Auxiliary Metering Inputs |
| |
|
Setting |
Range |
Resolution |
Accuracy |
|
Vdc/ac rms |
300.0V |
0.1V |
±0.7% rdg±5d |
|
Idc/ac rms |
5.000A CF<3
9.999A CF<1.5 |
0.001A |
±0.7% rdg±5d |
|
Phase |
-179.9° to +180.0° |
0.1° |
±3° |
|
Frequency |
40-100Hz |
0.01Hz |
±0.02% rdg±2d |
|
Protection: fuse on current input |
| |
| Timing System |
| |
| Range |
0-999.999s/9999.99s/99999.9s
autoranging |
| Resolution |
1/10/100ms |
| Accuracy |
0.01%rdg+2d (+4d current operated
mode) |
| Contact o/c |
24V |
| Contact s/c |
20mA |
| Vdc |
24—240V |
|
| |
| Each contact circuit will auto-select for
normally open or normally closed contacts. A DC voltage of 24—240Vdc
may also be used to trigger either timer channel. Contact state is
shown by an LED. |
| |
| Supply Requirements |
| |
| 115V/230V ±10% auto-selecting 50/60Hz 1ph
2300VA max. |
| |
| Safety |
| |
|
An earth terminal is
provided for connection to a local earth. The unit is designed to
comply with BSEN61010 and is CE marked. |
| |
| Temperature Range |
| |
Storage -20°C to 60°C
Operating0°C to 45°C |
| |
| Dimensions |
| |
| |
Dimensions |
Weight |
| Standard unit |
380mm x 314mm x 221mm |
19.5kg |
| Pelicase unit |
560mm x 456mm x 265mm |
23.3kg |
|
| |
| Cases and Options |
| |
| Unit
with Standard Aluminium Case |
Pelicase option
|
Pelicase option
|
 |
 |
 |
|