Auxiliary Output
A single phase isolated auxiliary output is provided on the
50A-3PH with a maximum rating of 33VA. |
|
Range |
Maximum VA |
Output Current |
|
|
110V |
33VA |
300mA |
|
|
230V |
33VA |
150mA |
|
|
Lead Set specifications
The 50A-3PH is supplied with a lead set including the
following items:
6 x 3m 4mm2 output leads terminated in 4mm plugs.
2 x 3m 2.5mm2 auxiliary leads terminated in 4mm plugs
2 x 0.5m 2.5mm2 link leads terminated in 4mm plugs
The lead set is supplied in a plastic carrying case.
Supply Requirements
The supply voltage requirements for the unit must be specified
at the time of ordering. The unit is available for
operation from either a 400V 4 wire supply or 220V 3 wire
supply. The optional delta-star supply transformer allows the
400V unit to operate from other supply voltages.
|
|
Supply option 1: |
400V-10%+14% |
50/60Hz |
3ph |
1000VA |
|
|
Supply option 2: |
220V -6%+14% |
50/60Hz |
3ph |
1000VA |
|
|
Timing System |
| Range |
0 to 999.999s |
|
| Resolution |
1ms |
|
| Accuracy |
0.1% of reading +2
digits |
(+4 digits in current operated
mode) |
|
The contact circuit has an open circuit voltage of 24Vdc and a
short circuit current of 10mA. Each contact circuit will
auto-select for normally open or normally closed contacts. A
DC voltage of 24-240Vdc may also be used to trigger either
timer channel.
The following functions are provided: |
|
Mode |
Timer Start |
Timer Stop |
|
|
Internal Start |
Press 'ON' |
Contact 1 |
|
|
Single Contact |
Contact 1 |
Contact 1 |
|
|
Dual Contact |
Contact 1 |
Contact 2 |
|
|
Current Operated |
Current >10% of range |
Current< 10% of range |
|
|
Pulse mode is used for setting the
current level in devices sensitive to heating, and allows
current to be injected for 600ms and the current recorded.
Current operated mode operates on one
output phase (selectable). The output is automatically
switched off at the end of the test to safeguard the relay
under test.
RS232 Communication Port
An RS232 port is provided to allow connection of a printer.
Protection and Safety
The unit is protected by electronic over current and duty
cycle trips on the outputs, thermal monitoring on the
power components, and fuses on the input and regulator. An
earth terminal is provided for connection to a local
earth.
The unit is designed to comply with BSEN61010, and is CE
marked.
Temperature Range |
|
Storage Temperature |
|
-20°C
to 60°C |
|
Operating Temperature |
|
0°C
to 45°C |
|
Dimensions |
380mm x 314mm x 221mm
Weight
21kg (including handle and corner protectors)
Accessories Supplied with
Instrument
Operating manual, output lead
set, mains lead, spare fuse set.
Optional Accessories
Delta-Star supply
transformer, filter unit and printer.
Optional Delta-Star
Supply Transformer
The optional delta-star supply transformer allows operation
from 115V, 230V, 400V, and 440V 3 wire supplies, is selected
by a switch on the unit. An auxiliary single phase output is
also provided to supply power to a DVS3 voltage source. |
| Input |
115V, 230V,400V, 440V ±10%
|
|
| |
3-wire, 1500VA maximum |
|
| 3Φ Output |
400V 4 wire 600VA 1 minute
'ON' 15 minutes 'OFF' |
|
| 1Φ
Output |
230V 300VA
5 minutes 'ON' 15 minutes 'OFF' |
|
|