High Current Test System

750ADM/750ADM-H  Current Injection Systems

 

You are here:

Models 750ADM/750ADM-H Specifications

Navigation:

High Current Tester Index >  T&R Test Equipment Index > 750ADM/750ADM-H Features

 
750ADM/750ADM-H Specifications
 

Main Output
The main output on the unit has two taps, allowing the selection  of output voltages up to 16V and output currents up to 750A. The unit operates at slightly reduced ratings when operating from a 115V supply.
 

750Amp Output 750ADM 750ADM-H  
  115V 230V 115V 230V  
Open Circuit Voltage 2.5V 3.5V 3.5V 5.0V  
Voltage at 500A 1.9V 2.8V 2.8V 4V  
Continuous Current 125A 125A 125A 125A  
5 minutes 'ON' 250A 250A 250A 250A  
1 minute 'ON' 500A 500A 440A 500A  
Maximum Current 750A 750A 500A 750A  
Maximum Current 'ON' time 10s 30s 10s 20s  
           
40Amp Output 750ADM 750ADM-H  
  115V 230V 115V 230V  
Open Circuit Voltage 10V 16V 10V 16V  
Full Load Voltage 7.5V 10V 7.5V 10V  
Continuous Current 10A 10A 10A 10A  
1 minute 'ON' 40A 40A 40A 40A  
Maximum current 'ON' time 10s 30s 10s 20s  

Metering

The output is metered by a digital true RMS system with a memory ammeter - whenever the output is switched off, the current reading is held on the display.
Range Resolution Trip Current Accuracy  
20.00A 0.01A 21A ±0.5% of reading +5digits  
50.00A 0.01A 53A ±0.5% of reading +5digits  
200.0A 0.1A 210A ±0.5% of reading +5digits  
750A 1A 788A ±0.5% of reading +2digits  

A current trip is automatically set to 105% of full scale of the selected metering range to protect the device under test.

Timing System
Range 0.999.999s  
Resolution 1ms  
Accuracy ±0.01% of reading +2 digits (all modes except current operated)
  ±0.01% of reading +4 digits (current operated mode)

The contact circuits have an open circuit voltage of 24Vdc and a short circuit current of 20mA. Each contact circuit will auto-select for normally open or normally closed contacts. A DC voltage of 24-240Vdc may also be used to trigger either timer channel.
 
Mode Timer Start Timer Stop  
OFF Timer inactive Timer inactive  
Internal Start Press 'ON' Contact 1  
Single Contact Contact 1 Contact 1  
Dual Contact Contact 1 Contact 2  
Current Operated Current > 20% of range Current < 20% of range  
The output is automatically switched off at the end of the test to safeguard the relay under test.

RS-232 Communication Port
An RS232 port is provided to allow connection to a PC or a printer. ADMlog software is available as an optional extra to log results from the 750ADM and transfer them to a spreadsheet.

T&R Link
The T&R Link allows a T&R DVS3 mk2 voltage source to phase lock to the 750ADM current.

Supply Requirements
Supply 750ADM 750ADM-H  
115V ±10% 50/60Hz 1ph 2000VA maximum 1900VA maximum  
230V ±10% 50/60Hz 1ph 3000VA maximum 3900VA maximum  

The correct voltage range is automatically selected by the unit.
 
Temperature Range
Storage -20°C to 60°C
Operating 0°C to 45°C

Dimensions
340mm x 230mm x 230mm
Weight
21.5kg (750ADM),   23.5kg (750ADM-H)

Accessories Supplied with Units
Operating manual, mains lead, spare fuse set and carrying strap.

Optional 750ADM-AL Lead Set specifications
A range of output leads are available to complement the 750ADM. The standard 750ADM-AL lead set is 3m long,
recommended for use with a 230V Supply. A 1.5m lead set is also available, and is recommended when operating from a 115V supply. The leads consist of double insulated 95mm2 welding cable terminated in Dinse high current connectors at the 750ADM end and high current welding clamps at the load end.
 

T & R Test Equipment Double Insulated 95mm2 welding cable terminated in Dinse high current connectors at the 750ADM end and high current welding clamps at the load end.

Low current timer leads are also included with the lead set:

 

T & R Test Equipment Double Insulated 95mm2 welding cable terminated in Dinse high current connectors at the 750ADM end and high current welding clamps at the load end.

The standard 3m lead set weighs 9.8kg including high current
leads, timer leads and case.
 

Protection and Safety
The unit is protected by electronic over current and thermal trips on the outputs, and circuit breakers on the input and power circuit. An earth terminal is provided for connection to a local earth. The unit is designed to comply with BSEN61010, and is CE marked.

 

Accles & Shelvoke| Aqua-Tronics | Arbiter Systems | Clare Instruments | Haefely | Hipotronics | HV Diagnostics | Megger | Tettex | T & R Equipment  
Home Page | High Current | High Voltage | Cable Fault | Oil Test | Transformer | Insulation 

 
Macey's Electrical Pty Ltd
ABN 17 344 601 468
Unit 1, 49 Gavenlock Road,  Tuggerah,  Sydney  Australia  2259

 

Phone: 61 2 4353 4644 

Mobile: 0412 678 925

Fax: 61 2 4353 4424

mailto:sales@maceyselectrical.com.au
 

Copyright: 1978 to 2007                                                   Last update: Friday, 03 October 2008