KV5-100
Specifications
Output
The main output on the unit is variable between 0 and 5kVac. The
output is rated at 50mA continuously and 100mA for 5 minutes
followed by an off time of 15 minutes.
Metering
The output voltage is metered by an analogue instrument with 0-3kV
and 0-6kV scaling.
|
Range |
Accuracy |
|
|
3kV |
±1.5% of FS |
|
|
6kV |
±1.5% of FS |
|
The output current is metered by a dual range analogue instrument
with 0-10mA and 0-100mA ranges. The current trip may be set to
20-120% of the selected range in 20% steps.
|
Range |
Trip Current |
Accuracy |
|
|
10mA |
2 to 12 mA |
±3% of FS |
|
|
100mA |
20 to 120mA |
±3% of FS |
|
Overload Protection
Two overload protection circuits are provided on the KV5- 100. The
first is user selectable, and allows trip currents
between 2mA and 120mA to be set. The second trip circuit monitors
the primary of the HV transformer, and
operates at approximately 130% or rated output current. An
overload/trip condition is indicated by an illuminated
push button and an audible alarm.
Burn
Out Circuit
When in circuit, the maximum short circuit current is limited to
100mA. When out of circuit, the maximum short
circuit current is approximately 2A.
Interlock Circuits
The unit has a zero volt interlock that prevents the output being
energized unless the output control is in the zero
position. An external interlock connection is also provided,
allowing the fitting of external emergency off buttons and test
cage door interlocks.
Supply
Requirements
115V±10% 50/60Hz 1ph 600VA max
240V±10% 50/60Hz 1ph 600VA max
Protection and Safety
The unit is protected by electronic over current and duty cycle
trips on the outputs and a fuse on the mains. An
earth terminal is provided for connection to a local earth. The
unit is designed to comply with BSEN61010, and is
CE marked.
Temperature Range
Storage -20°C to 60°C, Operating 0°C to 45°C
Dimensions and Weight
364mm (H) x 147mm (W) x 262mm (D) Weight: 16.5kg
Accessories Supplied
with Unit
Operating manual, mains lead, spare fuse. |