Metering Range |
Resolution |
Accuracy |
Current Trip |
|
|
10.00A |
0.01A |
±0.05% of rdg+2 digits |
10.5A |
|
|
20.00A |
0.01A |
±0.05% of rdg+2 digits |
21A |
|
|
100.0A |
0.1A |
±0.05% of rdg+2 digits |
100A |
|
| The output is
protected by electronic trips and a circuit breaker. |
Supply Requirements
230V±10%, 45-65Hz 1ph 11.5kVA max (23kVA overload for 2s)
Control Unit
Standard Accessories Supplied
Mains lead (5m), loading unit power and metering leads (5m),
operating manual and spare fuses.
Dimensions
PCU1-SP/PCU1-LT 450(W) x 275(H) x 305mm(D)
NLU2000/NLU5000 450(W) x 275(H) x 370mm(D) incl. terminals
Weight
PCU1-LT 29kg, PCU1-SP 35kg.
Temperature Range
Storage -20°C to 60°C, Operating 0°C to 45°C
Protection and
Safety
The PCU1 series and loading units are CE marked and are
designed to meet the requirements of BS EN61010. The system is
protected by electronic trips on the outputs, circuit breakers
on the mains input, control unit output and secondary
injection output (PCU1-SP only). The unit also has a duty
cycle trip on the loading unit output and the loading unit has
comprehensive thermal protection.
Optional Loading
Unit Specifications
Two loading units are available to provide a range of output
currents suitable for different primary injection tasks. Each
loading unit has three output taps allowing current injection
into a wide range of loads of differing impedances. Optional
output lead sets are also available.
NLU5000 Loading Unit
Intermittent Ratings |